HOME    About this site    mypage    Japanese    library    university    Feedback

University of the Ryukyus Repository >
Faculty of Science >
Bulletin >
Bulletin of the College of Science, University of the Ryukyus >
No.84 (2007/10) >

 
Title :蛍光X線分析装置(XRF)による1:5希釈ガラスビードを用いた全岩主成分・微量成分元素の定量分析
Title alternative :Analytical accuracy and precision of major and trace elements for bulk rocks using a 1:5 dilution glass bead by XRF
Authors :新城, 竜一
宮本, 正雪
Authors alternative :Shinjo, Ryuichi
Miyamoto, Masayuki
Issue Date :Oct-2007
Abstract :Analytical results of major and trace elements(Ba, Co, Cr, Nb, Ni, Rb, Sr, V, Y, and Zr) for bulk rocks by XRF using 1:5 dilution glass bead were evaluated. Analytical conditions and protocol are also reported. The results show relatively good accuracy and precision for major and trace elements. This method could be suitable to basalts to andesites, with error <$\pm$3% for major elements and <$\pm$10% for trace elements.
Type Local :紀要論文
ISSN :0286-9640
Publisher :琉球大学理学部
URI :http://hdl.handle.net/20.500.12000/5367
Citation :琉球大学理学部紀要 = Bulletin of the College of Science. University of the Ryukyus no.84 p.5 -13
Appears in Collections:No.84 (2007/10)

Files in This Item:

File Description SizeFormat
No84p5.pdf652KbAdobe PDFView/Open